USG-400
ISSUE1 8-13
Troubleshooting
Irregularity Description
Possible cause
Remedial Actions
The output volume of
THUNDERBEAT can not
be adjusted.
The output volume of
THUNDERBEAT can not be
adjusted with the ultrasonic
generator.
Adjust the output volume of
THUNDERBEAT with the compatible
electrosurgical generator.
The ultrasonic generator
cannot be turned OFF.
Ultrasonic generator malfunction.
Remove the THUNDERBEAT and/or
SONICBEAT from the body cavity and press
the power switch to OFF. Contact Olympus.
The settings are erroneous.
Correct the settings of the compatible high
flow insufflation unit.
The communication cable is not
connected.
Connect the communication
cable. ”Automatic mist & smoke
evacuationsystem/function (when using
the compatible high flow insufflation unit)”
The connection of the
communication cable is
erroneous.
Reconnect the communication cable.
“Automatic mist & smoke evacuation
system/function (when using the
compatible high flow insufflation unit)”.
Compatible high flow insufflation
unit malfunction.
Contact Olympus.
Automatic mist & smoke
Evacuation system/function
does not work.
Ultrasonic generator,
communication cable or adapter
for UHI-2/3 malfunction
Remove the THUNDERBEAT and/or
SONICBEAT from the body cavity and
press the power switch to OFF. Contact
Olympus.
Operation is not appropriate.
Operate the instrument properly referring to
Chapter 4, “Operation” in the instruction
manual for THUNDERBEAT or SONICBEAT.
The treated site is not appropriate.
Treat the adequate site referring to
Chapter 4, “Operation” in the instruction
manual for THUNDERBEAT or
SONICBEAT.
Carbonized tissue is present on the
grasping section or the probe tip.
Remove the carbonized tissue from the
grasping section or the probe tip with sterile
gauze or soft brush.
The output level setting is not
appropriate.
Set the adequate output level referring to
Chapter 3, “Preparation and Inspection” in
the instruction manual for THUNDERBEAT
or SONICBEAT.
Deterioration of the transducer.
Confirm either whether one year has
passed since the transducer was delivered
or the use of the transducer has exceeded
100 cases. If it is applied to at least one of
them, replace the transducer with a new
one.
Cutting/sealing/
coagulating does not work
properly.
Not activating output.
Refer to “Energy output cannot be
enerated” of “What to do when no error
code is displayed” in Section 8.3,
“Troubleshooting guide”.
Summary of Contents for USG-400
Page 1: ...ISSUE 1 2011 11 MAINTENANCE MANUAL USG 400...
Page 2: ......
Page 6: ......
Page 10: ...USG 400 Product Outline 1 4 ISSUE1 4 System chart...
Page 11: ...USG 400 ISSUE1 2 1 Specifications Chapter 2 Specifications 1 Specifications 2 2...
Page 14: ......
Page 34: ......
Page 38: ......
Page 43: ...USG 400 ISSUE1 5 5 Inspection Figure 5 3 Figure 5 4...
Page 50: ...USG 400 Inspection 5 12 ISSUE1 Figure 5 12...
Page 59: ...USG 400 ISSUE1 6 1 Safety check Chapter 6 Safety check 1 Check procedures 6 2 2 Check Card 6 6...
Page 65: ...USG 400 ISSUE1 7 1 Block Diagram Chapter 7 Block Diagram 1 Block Diagram 7 2...
Page 89: ......