USG-400
ISSUE1 8-19
Troubleshooting
Error Code
Irregularity
Description
Possible cause
Remedial Actions
Output is generated while a
metallic object is grasped.
1. Confirm no metal object is grasped
between jaw and probe, then
confirm activation.
Output in body fluids.
2. Remove excess fluids in the body.
U508
SEAL & CUT short
circuit error
Grasping section wear.
3. If the error window is still displayed,
replace the THUNDERBEAT or
SONICBEAT instrument.
Hard tissue (e.g., bone) is
caught.
1. Do not grasp the hard tissue.
Excessive load is applied to the
distal end.
2. Reduce force applied to distal end
of the probe.
Connection of the transducer
with THUNDERBEAT or
SONICBEAT instrument is loose.
3. Check if the THUNDERBEAT or
SONICBEAT instrument and
transducer are connected firmly. If
not, connect them firmly.
Transducer plug is not
connected to the
THUNDERBEAT or SONICBEAT
socket properly.
4. Confirm that the transducer plug is
connected to the THUNDERBEAT
or SONICBEAT socket on the front
panel of the ultrasonic generator. If
not, connect it properly.
Transducer malfunction.
5. Replace the transducer.
Probe damage or malfunction.
6. Replace the THUNDERBEAT or
SONICBEAT instrument.
U509
Ultrasonic level error
Ultrasonic generator malfunction.
7. Contact Olympus.
Summary of Contents for USG-400
Page 1: ...ISSUE 1 2011 11 MAINTENANCE MANUAL USG 400...
Page 2: ......
Page 6: ......
Page 10: ...USG 400 Product Outline 1 4 ISSUE1 4 System chart...
Page 11: ...USG 400 ISSUE1 2 1 Specifications Chapter 2 Specifications 1 Specifications 2 2...
Page 14: ......
Page 34: ......
Page 38: ......
Page 43: ...USG 400 ISSUE1 5 5 Inspection Figure 5 3 Figure 5 4...
Page 50: ...USG 400 Inspection 5 12 ISSUE1 Figure 5 12...
Page 59: ...USG 400 ISSUE1 6 1 Safety check Chapter 6 Safety check 1 Check procedures 6 2 2 Check Card 6 6...
Page 65: ...USG 400 ISSUE1 7 1 Block Diagram Chapter 7 Block Diagram 1 Block Diagram 7 2...
Page 89: ......