10-004209-01EN [Q7750159], Rev. 2, May 2018
Troubleshooting
209
Table 10 Troubleshooting table
Problem
Possible cause
Solution
Handheld controller
display does not activate
Input power
requirements not met.
Ensure input power
meets requirements (see
“Power Requirements”
on page 219).
Handheld controller not
plugged into power
controller.
Plug handheld controller
into power controller.
Ensure connectors are
dry, clean and connector
pins are not bent.
Umbilical cable not
properly connected.
Check umbilical cable
connections at both ends.
Ensure connectors are
dry and clean and
connector pins are not
bent.
SteerROVER system not
started.
Start the SteerROVER
system (see “System
Start-Up” on page 57).
Damaged components in
the handheld controller,
scanner, power
controller, or cabling.
Contact Olympus
technical support.
Handheld controller
display is activated, yet
scanner does not drive
Handheld controller is
not in correct mode for
driving.
See “Operation” on
page 57 for additional
details.
Damaged components in
the handheld controller,
scanner, power
controller, or cabling.
Contact Olympus
technical support.
Summary of Contents for SteerROVER
Page 8: ...10 004209 01EN Q7750159 Rev 2 May 2018 Table of Contents viii...
Page 10: ...10 004209 01EN Q7750159 Rev 2 May 2018 List of Abbreviations x...
Page 18: ...10 004209 01EN Q7750159 Rev 2 May 2018 Labels and Symbols 8...
Page 38: ...10 004209 01EN Q7750159 Rev 2 May 2018 Introduction 28...
Page 52: ...10 004209 01EN Q7750159 Rev 2 May 2018 Chapter 1 42...
Page 66: ...10 004209 01EN Q7750159 Rev 2 May 2018 Chapter 2 56...
Page 208: ...10 004209 01EN Q7750159 Rev 2 May 2018 Chapter 4 198...
Page 212: ...10 004209 01EN Q7750159 Rev 2 May 2018 Chapter 5 202...
Page 222: ...10 004209 01EN Q7750159 Rev 2 May 2018 Chapter 6 212...
Page 224: ...10 004209 01EN Q7750159 Rev 2 May 2018 Chapter 7 214...