![Nidek Medical ICE mini Operator'S Manual Download Page 40](http://html1.mh-extra.com/html/nidek-medical/ice-mini/ice-mini_operators-manual_1688064040.webp)
30
OPERATING PROCEDURES
:
Operating Procedures
2.2.5
Blocking with frame change holder
Use the provided holder when the lens can not be held with the three pins on the lens table because the
lens diameter is small as for when lenses are being reused in different frames.
1
Place the frame change holder on the center
of the lens table.
The orientation of the frame change holder is
as shown to the right.
2
Place the lens on the frame change holder and block it.
Block the lens while holding it by hand without the lens clamp.
CAUTION
• Confirm that the lens is held securely with the three pins on the frame change holder.
The lens can not be blocked correctly when the lens is held up by contact with one or
more of the pins on the lens table.
In such cases, pull out the pins on the lens table before using the frame change holder.
• Hold the lens so that the hand is not in contact with the blocking arm and/or lens cup.
Frame change
holder