Specifications
N-600x Operator’s Manual
179
Battery
The battery provides at least 7 hours of battery life when new and fully
charged with no alarms, no serial data, no analog output, no nurse call
output, with backlight on while using an SRC-MAX set at 200 bpm, high
light and low modulation.
Type
Lead acid
Voltage
6 Volts DC
Recharge
• 8 hours with N-600x turned
off
• 12 hours with N-600x turned
on
Shelf Life
• 4 months, new fully-charged
battery (when monitor is
placed in "Shelf-mode" by
qualified service personnel
using the procedures
indicated in the
N-600x Service Manual
)
• After 4 months storage in
"Shelf-mode", the N-600x
will run for 33% of stated
battery life
Complies With
91/157/EEC
O
XI
M
AX
Sensors
Wavelength
and Power
Nellcor pulse oximetry sensors contain LEDs that emit
red light at a wavelength of approximately 660 nm and
infrared light at a wavelength of approximately 900 nm.
The total optical output power of the sensor LEDs is less
than 15 mW. This information may be useful to
clinicians, such as those performing photodynamic
therapy.
Summary of Contents for OXIMAX N-600X Series
Page 1: ......
Page 8: ...Contents vi...
Page 12: ...Safety Information 4 N 600x Operator s Manual...
Page 80: ...Using the Monitor 72 N 600x Operator s Manual...
Page 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Page 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Page 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Page 128: ...Using the Data Port 120 N 600x Operator s Manual...
Page 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Page 146: ...Performance Considerations 138 N 600x Operator s Manual...
Page 168: ...Troubleshooting 160 N 600x Operator s Manual...
Page 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Page 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Page 204: ...Specifications 196 N 600x Operator s Manual...
Page 210: ...Index 202 N 600x Operator s Manual...
Page 211: ......