Printing Monitor Trend Data
108
N-600x Operator’s Manual
Time
The Time column displays the value of the N-600x real-time clock.
Patient Data
Patient information is highlighted in the display above. Parameter
values are displayed directly beneath the heading for each parameter.
In this example, the %SpO
2
is 100 and the pulse rate is 190 beats per
minute. The “*” next to the 190 indicates that 190 beats per minute is
outside of the alarm limits, indicated in the top row, for pulse rate. If
no data for a parameter is available, three dashes [- - -] display.
PA represents the pulse amplitude value, in which the number can
range from 0 to 254. There are no alarm parameters for this value. It
can be used for trending information as an indication of a change in
pulse volume, relative pulse strength, or circulation.
Operating Status
Summary of Contents for OXIMAX N-600X Series
Page 1: ......
Page 8: ...Contents vi...
Page 12: ...Safety Information 4 N 600x Operator s Manual...
Page 80: ...Using the Monitor 72 N 600x Operator s Manual...
Page 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Page 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Page 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Page 128: ...Using the Data Port 120 N 600x Operator s Manual...
Page 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Page 146: ...Performance Considerations 138 N 600x Operator s Manual...
Page 168: ...Troubleshooting 160 N 600x Operator s Manual...
Page 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Page 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Page 204: ...Specifications 196 N 600x Operator s Manual...
Page 210: ...Index 202 N 600x Operator s Manual...
Page 211: ......