Specifications
N-600x Operator’s Manual
189
Radiated RF
IEC
61000
-
4
-
3
3 V/m
80 MHz
800 MHz
3 V/m
distance
= 1.2
√
Power
80 MHz to 800 MHz
3 V/m
800 MHz
2.5 GHz
3 V/m
distance
= 2.3
√
Power
800 MHz to 2.5 GHz
Table 19: Electromagnetic Immunity, RF Portable Equipment
For portable and mobile communication equipment. The N
-
600x is
suitable for use in the specified electromagnetic environment. The
customer and/or user of the N
-
600x should assure that it is used in an
electromagnetic environment as described below:
Immunity
Test
IEC
60601
-
1
-
2
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
Note
: Field strengths from fixed transmitters, such as base stations
for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast, and TV broadcast
cannot be predicted theoretically with survey accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured
field strength in the location in which the N
-
600x is used exceeds
the applicable RF compliance level above, the N
-
600x should be
observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as
re
-
orienting or relocating the N
-
600x.
Note
: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Interference may occur in the vicinity of equipment marked with the
following symbol:
Summary of Contents for OXIMAX N-600X Series
Page 1: ......
Page 8: ...Contents vi...
Page 12: ...Safety Information 4 N 600x Operator s Manual...
Page 80: ...Using the Monitor 72 N 600x Operator s Manual...
Page 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Page 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Page 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Page 128: ...Using the Data Port 120 N 600x Operator s Manual...
Page 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Page 146: ...Performance Considerations 138 N 600x Operator s Manual...
Page 168: ...Troubleshooting 160 N 600x Operator s Manual...
Page 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Page 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Page 204: ...Specifications 196 N 600x Operator s Manual...
Page 210: ...Index 202 N 600x Operator s Manual...
Page 211: ......