Using the Monitor
60
N-600x Operator’s Manual
Selecting the Standby Mode
The standby mode enables the monitor to retain the alarm limit
settings that are in effect while monitoring a patient. The monitor
must be connected to an AC power source to enter the standby mode.
Normally, the standby mode setting is used when a patient has to
temporarily leave the monitor.
1. Verify the N-600x is monitoring a patient and the alarm limits are
configured to the patient being monitored.
2. Disconnect the sensor from the monitor.
3. Press the ALARM SILENCE button to silence the audible
alarms.
4. Press the ALARM SILENCE button to disable the alarm
messages.
The monitor is now in the standby mode. Reconnect the sensor to the
monitor and the patient to return to normal monitoring.
Summary of Contents for OXIMAX N-600X Series
Page 1: ......
Page 8: ...Contents vi...
Page 12: ...Safety Information 4 N 600x Operator s Manual...
Page 80: ...Using the Monitor 72 N 600x Operator s Manual...
Page 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Page 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Page 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Page 128: ...Using the Data Port 120 N 600x Operator s Manual...
Page 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Page 146: ...Performance Considerations 138 N 600x Operator s Manual...
Page 168: ...Troubleshooting 160 N 600x Operator s Manual...
Page 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Page 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Page 204: ...Specifications 196 N 600x Operator s Manual...
Page 210: ...Index 202 N 600x Operator s Manual...
Page 211: ......