Specifications
178
N-600x Operator’s Manual
Electrical
1
Adult specifications are shown for
O
XI
M
AX
MAX
-
A and MAX
-
N
sensors with the N
-
600x. Neonate specifications are shown for
O
XI
M
AX
MAX
-
N sensors with the N
-
600x. Saturation accuracy varies by sensor
type. Refer to the Sensor Accuracy Grid shipped with the monitor.
The Sensor Accuracy Grid is available on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/ProductManuals.html
2
Specification applies to N-600x monitor performance.
Reading accuracy in the presence of low perfusion (detected IR pulse
modulation amplitude 0.03% - 1.5%) was validated using signals
supplied by a patient simulator. SpO2 and pulse rate values were varied
across the monitoring range over a range of weak signal conditions and
compared to the known true saturation and pulse rate of the input
signals.
Display Update Interval
2 seconds
Accuracy Tolerance
Instrument
Power
Requirements
Rated at 100 to 120 volts AC (nominal 120 VAC) or
220 to 240 volts AC (nominal 230 VAC), 20 volt/
amps to be compliant with IEC 60601
-
1 sub
-
clause
10.2.2.
Fuses
Quantity 2, 0.5 A, 250 volts, slow
-
blow,
IEC (5 x 20 mm).
Summary of Contents for OXIMAX N-600X Series
Page 1: ......
Page 8: ...Contents vi...
Page 12: ...Safety Information 4 N 600x Operator s Manual...
Page 80: ...Using the Monitor 72 N 600x Operator s Manual...
Page 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Page 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Page 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Page 128: ...Using the Data Port 120 N 600x Operator s Manual...
Page 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Page 146: ...Performance Considerations 138 N 600x Operator s Manual...
Page 168: ...Troubleshooting 160 N 600x Operator s Manual...
Page 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Page 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Page 204: ...Specifications 196 N 600x Operator s Manual...
Page 210: ...Index 202 N 600x Operator s Manual...
Page 211: ......