Primary Speaker Failure
The N-600x may detect a failure of the primary speaker and sound a
high-pitched, slow-pulsing piezo tone. A primary speaker failure
message displays as shown below.
1. Press HELP to continue. The following message displays.
2. Press BACK to display the speaker failure message again. The
message cannot be cleared.
3. Press the ALARM SILENCE button to silence the slow-pulsing
piezo tone.
Note:
Once the monitor is silenced, the N-600x sounds a piezo tone
every three minutes as a reminder of the primary speaker
failure condition. The N-600x also sounds the piezo tone to
annunciate low, medium and high priority alarms during this
time. If an N-600x monitor is reporting a primary speaker
failure is powered off, it cannot be powered on again.
HELP
100
250
PRIMARY SPKR FAILURE:
NOTIFY SVC PERSONNEL.
PRESS HELP
BACK
100
250
HELP SPEAKER FAILURE
Note: Once this
monitor is powered
off, it cannot be
powered on again.
Summary of Contents for OXIMAX N-600X Series
Page 1: ......
Page 8: ...Contents vi...
Page 12: ...Safety Information 4 N 600x Operator s Manual...
Page 80: ...Using the Monitor 72 N 600x Operator s Manual...
Page 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Page 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Page 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Page 128: ...Using the Data Port 120 N 600x Operator s Manual...
Page 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Page 146: ...Performance Considerations 138 N 600x Operator s Manual...
Page 168: ...Troubleshooting 160 N 600x Operator s Manual...
Page 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Page 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Page 204: ...Specifications 196 N 600x Operator s Manual...
Page 210: ...Index 202 N 600x Operator s Manual...
Page 211: ......