Troubleshooting
146
N-600x Operator’s Manual
Table 12: Error Codes
Error
Code
Error
Message
Action
80
DEFAULTS
LOST
The current power
-
on default settings
have been lost and returned to factory
defaults. Qualified service personnel
can use the
N-600x Service Manual
to
restore the desired power
-
on default
settings.
81
SETTINGS
LOST
The current settings (for example, alarm
limits, alarm and pulse beep volumes,
alarm silence duration) have been lost
and returned to power
-
on defaults. Turn
the monitor off and back on again. If it
is necessary to have settings different
from the power
-
on default settings, turn
the monitor off and back on again, and
reenter the desired settings.
82
CLOCK
SETTING
LOST
The date and time settings have been
lost. Reenter the date and time.
Recharge or replace the battery.
515, 518,
534, 535,
569
N-600x Boot
Version x.x.x.x
The application software is missing or
corrupt. Notify a qualified service
personnel.
529, 729
LOW
BATTERY
The battery is discharged to a critically
low level. The monitor will shutdown
after 10 seconds.
Verify the SUPPLY VOLTAGE
SELECTOR switch on the rear panel is
set to the proper voltage.
Connect the monitor to AC power and
turn back on. A warning message is
displayed and a low priority audible
alarm sounded. These must be
acknowledged (by pressing the alarm
silence key twice) before the monitor
can be used for patient monitoring.
Summary of Contents for OXIMAX N-600X Series
Page 1: ......
Page 8: ...Contents vi...
Page 12: ...Safety Information 4 N 600x Operator s Manual...
Page 80: ...Using the Monitor 72 N 600x Operator s Manual...
Page 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Page 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Page 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Page 128: ...Using the Data Port 120 N 600x Operator s Manual...
Page 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Page 146: ...Performance Considerations 138 N 600x Operator s Manual...
Page 168: ...Troubleshooting 160 N 600x Operator s Manual...
Page 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Page 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Page 204: ...Specifications 196 N 600x Operator s Manual...
Page 210: ...Index 202 N 600x Operator s Manual...
Page 211: ......