Using the Data Port
N-600x Operator’s Manual
117
The nurse call function needs to be tested after it has been set up in
your facility. The nurse call feature should be tested whenever setting
up the N-600x pulse oximeter in a location that uses nurse call. If an
attached
O
XI
M
AX
sensor is not connected to a patient, the monitor
display reads zeros and the monitor remains in the Pulse Search Mode
for 5 seconds, then the monitor displays [
- - -
] (3 dashes) in the
%SpO
2
and pulse rate display. One way to test the nurse call function
is to create an alarm condition (for example, sensor disconnect) and
verify that your facility's nurse call system is activated.
Setting Nurse Call RS-232 Polarity
The nurse call polarity can be set to a positive signal (NORM +) on a
monitor alarm condition or a negative signal (NORM -) on a monitor
alarm condition.
1. With the monitor in the normal monitoring mode, press the
SETUP softkey.
2. Press the NEXT softkey twice and then press the NCALL
softkey.
3. Press the NORM + softkey OR press the NORM -
softkey.
4. Press the EXIT softkey.
Summary of Contents for OXIMAX N-600X Series
Page 1: ......
Page 8: ...Contents vi...
Page 12: ...Safety Information 4 N 600x Operator s Manual...
Page 80: ...Using the Monitor 72 N 600x Operator s Manual...
Page 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Page 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Page 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Page 128: ...Using the Data Port 120 N 600x Operator s Manual...
Page 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Page 146: ...Performance Considerations 138 N 600x Operator s Manual...
Page 168: ...Troubleshooting 160 N 600x Operator s Manual...
Page 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Page 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Page 204: ...Specifications 196 N 600x Operator s Manual...
Page 210: ...Index 202 N 600x Operator s Manual...
Page 211: ......