Audible Alarm Settings and Service Functions
32
Operator's Menu Options
The menu items listed below are options that are available to the operator.
These options can be used to print or clear trend data, and to change the
configuration of data from the data port. Changes can be made in menu
options 1 through 5 while monitoring a patient. Menu items beyond 5 cannot
be accessed when a sensor is connected to the N-550. Unless changes are
saved as power-on default values, they will be lost when the N-550 is turned
off. When the N-550 is turned on again, it will begin to use the power-on
11
-
Alarm silence restrictions
0
Allow “OFF” as a choice for alarm silence
duration, reminder will be sounded
0 — Allows operator to select alarm silence Off, but
there will be a reminder tone every 3 minutes
1
Allow “OFF” as a choice for alarm silence duration,
no reminder will be sounded
1 — Alarm silence off, there will be no alarm
reminder
2
Do not allow “OFF” as a choice for alarm silence
duration
2 — Not allow alarm silence Off
12
-
Silence alarms (No alarm condition, both audio and
visual alarms)
0
On
1
Off
13
-
N-550 display turned on
0
On
1
Off
14
-
Alarm sound selection
1
Nellcor style
2
Alarms sound IAW IEC 60601-1-8
3
Alarms sound IAW EN 475
Table 2: N-550 Menu Selections (Continued)
Menu
Number
Sub-Menu
Number
Function
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...