Performance Verification
26
2. The N-550 pulse amplitude indicator will fill (10 bars) and stabilize and
illuminate 4 bars with each pulse beat.
3. The N-550 will display:
•
75 %SpO2 (test pass criteria is 73 to 77 %SpO
2
)
•
60 bpm (test pass criteria is 57 to 63 bpm)
•
alarm
•
pulse amplitude indicator - high level modulation
on page 22. The pulse amplitude indicator should
indicate high level modulation.
on page 23. The pulse amplitude indicator should
indicate high level modulation.
6. Press the SRC-MAX Light Level selection button. The SRC-MAX Light
Level
LED will light.
7. The N-550 pulse amplitude indicator will stabilize and illuminate 4 bars
with each pulse beat.
8. The N-550 will display:
•
75 %SpO2
•
60 bpm
•
alarm
•
low level modulation