Data Port Interface Protocol
N-550
91
5. Select baud rate by pressing Adjust Up or Adjust Down button (2400,
9600, or 19200 [default]).
6. Press the
SatSeconds
Alarm Limit button to set the selection.
Connecting To The Data Port
Data is transmitted in the RS-232 and RS-422 formats. RS-232 data can be
transmitted a maximum of 25 feet (7.62 meters). The RS-422 data can be
transmitted a maximum of 4,000 feet (1,219.2 meters). The pin outs for the
data port are listed in Table 13.
TXD+ and TXD- are the differential transmit data pair. RXD+ and RXD- are
the differential receive pair.
Table 13: Data Port Pinouts
Pin
Signal Name
1
RXD+ (RS-422 positive input)
2
RXD (RS
-
232 input)
3
TXD (RS
-
232 output)
4
TXD+ (RS-422 positive output) (-5 to -12 VDC with no audible
alarm, 5 to 12 VDC with audible alarm)
5
Signal Ground (isolated from Earth Ground)
6
NC (No connection)
7
Normally Open (N.O. with no audible alarm), dry contact for Nurse
Call
8
Normally Closed (N.C. with no audible alarm), dry contact for
Nurse Call)
9
RXD- (RS-422 negative output)
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...