Data Port Interface Protocol
96
Actual column headings are in the third row of the Column Heading line.
Patient data, from left to right, are the time that the chart was printed, the
current %SpO
2
value being measured, the current pulse rate in beats per
minute (bpm), the current Pulse Amplitude (PA), and the operating status of
the N-550.
Time Tag
Time Tag represents a real-time clock in: Day, Month, Year, and 24-hour
clock. The clock is maintained by either AC or battery power.
Patient Data
Patient data and the operating status of the N-550 are highlighted in the
patient data display. Parameter values are displayed directly beneath the
heading for each parameter. In this example, the %SpO
2
is 100 and the pulse
rate (PR) is 190 beats per minute. The asterisk (*) next to the 190 indicates
that 190 beats per minute is outside of the alarm limits, indicated at the
far-right end in the top row, for pulse rate. If no data for a parameter is
available, three dashes (- - -) will be displayed in the printout.
The number under PA is an indication of pulse amplitude. The number can
range from 0 to 254 and will typically range around 45. There are no alarm
parameters for this value. It can be used for trending information and indicates
a change in pulse volume, pulse strength, or circulation.
Operating Status
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...