Data Port Interface Protocol
N-550
97
The Status column indicates alarm conditions and operating status of the
N-550. The PH in this example indicates a Pulse Rate Upper alarm. The
Status column can have as many as four codes displayed in one line of data.
The status codes are listed in Table 14.
Note
:
A Sensor Disconnect will also cause three dashes (- - -) to be
displayed in the patient data section of the printout.
Table 14: Status Codes
Code
Meaning
AO
Alarm Off
AS
Alarm Silence
BU
Battery in Use
LB
Low Battery
LM
Loss of Pulse with Motion
LP
Loss of Pulse
MO
Motion
PH
Pulse Rate Upper Limit Alarm
PL
Pulse Rate Lower Limit Alarm
PS
Pulse Search
SD
Sensor Disconnect
SH
Saturation Upper Limit Alarm
SL
Saturation Lower Limit Alarm
---
No Data Available
*
Alarm Parameter Being Violated