Performance Verification
N-550
19
3. Press the Adjust Up button, while holding the Alarm Silence button, to
increase the alarm volume setting to a maximum value of “10.” Verify
that the volume increases.
4. Press the Adjust Down button, while holding the Alarm Silence button,
until a comfortable audio level is obtained.
5. Release the Alarm Silence button. The tone stops.
Pulse Tone Volume Control
1. When a valid pulse is acquired, press the Adjust Up button and verify that
the sound level of the beeping pulse tone increases.
2. Press the Adjust Down button and verify that the beeping pulse tone
decreases until it is no longer audible.
3. Press the Adjust Up button to return the beep volume to a comfortable
level.
Pulse Oximetry Functional Tests
These tests utilize the pulse oximetry functional tester (SRC-MAX) to verify
the performance of the N-550 pulse oximeter.
All of these tests should be done in sequence.