Data Port Interface Protocol
92
The pin layouts are illustrated in Figure 42. The conductive shell is used as
earth ground. An AMP connector is used to connect to the data port. Use
AMP connector (AMP P/N 747538-1), ferrule (AMP P/N 1-747579-2) and
compatible pins (AMP P/N 66570-2). The serial cable must have a braided
shield providing 100% coverage, such as Beldon cable (Beldon P/N 9616) or
equivalent. Connectors at both ends of the serial cable must have the shield
terminated to the full 360 degrees of the connector's metal shell.
CAUTION: Do not create sharp bends in the cable; this may tear or
break the shield.
Figure 42: Data Port Pin Layout
Pins 2, 3, and 5 provide data in RS
-
232 format.
Pins 1, 4, 9, and 12 provide data in RS-422 format.
10
Signal Ground
11
Nurse Call (RS-232 level output)
12
TXD- (RS-422 negative output) (-5 to -12 VDC with no audible
alarm, 5 to 12 VDC with audible alarm)
13
NC
14
NC
15
Nurse Call Common for Dry Contacts
Table 13: Data Port Pinouts (Continued)
Pin
Signal Name
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...