Audible Alarm Settings and Service Functions
42
5. Simultaneously press SpO
2
Alarm Limit button and Pulse Rate Alarm
Limit button for over three seconds. Menu Item 1 screen appears.
6. Press the Adjust Up button until menu Item 9 is displayed.
7. Press the
SatSeconds
Alarm Limit button to save all settings as the
institutional default settings.
8. The N-550 sounds three beeps, indicating that defaults are reset.