Cleaning and Maintenance
JuLI™ Stage - User Manual
61
IMPORTANT!
The well plate’s cover should be wetted by the cell culture media before monitoring.
This step is required to avoid condensation inside the well plate cover for monitoring.
IMPORTANT!
Avoid exposing the JuLI
TM
Stage
to UV light.
UV light may degrade components, including plastics. Damage from UV exposure
is not covered under the manufactu
rer’s warranty.
IMPORTANT!
Always wipe surfaces with ethanol-soaked paper towels.
☞
Note: The objectives are especially prone to damage by excessive ethanol.
Please always clean the objectives with ethanol-wetted objective cleaning paper.
Do not directly spray ethanol anywhere on the JuLI
TM
Stage.
IMPORTANT!
When the JuLI
TM
Stage is not used for long periods of time, it must be taken out from
the incubator to reduce humidity and put in a clean and stable place.