Installation
JuLI™ Stage - User Manual
17
IMPORTANT!
Use the cover if the instrument operates on the table to protect test samples from
external light.
4.3 Operating environment
Allow at least 10 cm (4 inches) free space at the back of the instrument to allow
for proper ventilation and prevent overheating of electronic components.
Set up the
JuLI™ Stage away from direct light sources, such as windows.
Ambient room lighting can enter the imaging path and affect the image quality.
Place the
JuLI™ Stage
on a level surface away from vibrations from other
pieces of equipment.
Operating temperature range: 5 - 40°C (41 - 104°F)
Relative humidity range: 20 - 95 %.
4.4 Cautions
IMPORTANT!
First use
: Before operating the
JuLI™ Stage
system for the first time, we strongly
recommend allowing the instrument to equilibrate to room temperature for
20 minutes
outside the incubator while the device is turned on.
If the
JuLI™ Stage is to be operated inside an incubator, an additional warm-up
time of
2 hours
is required to equilibrate it to 37 °C. During this time the
JuLI™
Stage must be switched on.
In
an incubator, the JuLI™ Stage and PC must be left turned on at all times.