Control Panel Operations
MTS Exceed™ 22 Series Product Information
Operation
45
[Impact]: Make the pendulum complete the impact process. Tap the
Impact
icon,
and the the pendulum hanging/release device moves so that the pendulum rotates
around the shaft under its weight to impact the specimen. When the impact is
complete, the pendulum is rotated counter-clockwise. When its angle approaches
zero, the brake device stops the pendulum, and the motor drives the pendulum to
rotate counter-clockwise until the induction switch is triggered and then releases
the pendulum. The pendulum is placed on the hanging rod under the influence of
its weight. The impact action is completed. It is equivalent to the
Impact
button
on the control panel and sliding door.
[Do not Use Pointer]: It refers to the test method that does not use pointer for
readings during the impact test. In such a test method, the equipment only
calculates air and bearing energy losses into absorption energy, excluding
frictional energy losses of pointers. Tap the
Do Not Use Pointer
icon to switch to
Use Pointer
test method. Under the
Use Pointer
test method, the system
incorporates air and bearing energy losses and pointer friction energy losses into
the calculation of absorption energy, as shown in the following figure:
Summary of Contents for Exceed 22 Series
Page 54: ...MTS Exceed 22 Series Product Information 54 Shortcut Operations Operation...
Page 58: ...MTS Exceed 22 Series Product Information 58 Troubleshooting...
Page 60: ...MTS Exceed 22 Series Product Information 60 Equipment Disassembly...
Page 67: ......
Page 68: ...m MTS Systems Corporation http www mts com en Global index asp ISO 9001 Certified QMS...