1. Touch Screen
Interfaces
1.1 Main Page
After the equipment is powered on, the touch screen displays the main page after
about 15s. You can select language on the main page. To ensure test data
accuracy, you should perform the clearing action each time the equipment is
powered on. As shown in the figure below:
[
中文
/EN]: Tap the
中文
/EN
icon to switch system language between Chinese
and English.
[Clear]: Tap the
Clear
icon, and the equipment automatically performs pendulum
collection and placement actions. After the pendulum is placed, it swings
automatically within a very small angle. When the pendulum stops swinging and
the value of
Current Angle
stops changing (as shown in the figure below), the
equipment automatically sets the angle to 0°. At this point, the clearing action is
completed. If you press any other buttons during clearing, no response will be
triggered. It is equivalent to the
Clear
button on the control panel.