Power Switch (I/O)
MTS Exceed™ 22 Series Product Information
Operation
37
Power Switch (I/O)
Power Switch (I/O)
Power switch of the main unit is located on the left of the electric control box in
the rear of the equipment. Turn the power switch to "I" to supply power to the
main unit and controller. When the power switch is turned on, LCD of the
controller panel is lighted. Turn the power switch to "O" to turn the power off.
Emergency stop
The tester is also equipped with
emergency stop
button that can cut off power of
the motor. This
button
can only be used for emergency purposes.
The emergency stop button should be checked regularly when the controller has
power but system is not running tests. The controller continuously monitors
electrical connections associated with the
emergency stop
button, and will alert
the user once it detects any errors. Press the
emergency stop
button to check
relevant conditions.
Press the
emergency stop
button, cut off power of the motor, and stop the test
program. Then, rotate the
emergency stop
button clockwise to release
emergency stop. In the case of emergencies, use this button to stop your tester.