MSA AUER
MSA
Maintenance and Service
SUPREMATouch
87
US
Examples
for MCI – Check 4 ... 20 mA
The sensor simulation module may be used only to the check and presetting and not to the cali-
bration.
for MPI – Check Catalytic Combustion Sensors
for MPI – Check Semiconductor Sensors
Sensor type:
PrimaX
Measuring gas:
Carbon monoxide
Zero gas:
Air
Reference gas:
Carbon monoxide
Ua at open switch [Normal operation]:
400 mV
Ua at closed switch [Alarm] :
1.9 V
Sensor type:
Series 47k
Measuring gas:
Methane
Zero gas:
Air
Reference gas:
Methane
Ux at open switch [Normal operation]:
0 mV
Ux at closed switch [Alarm] :
100 mV
Sensor type:
DF-8101
Measuring gas:
Acetone
Zero gas:
Air
Reference gas:
Acetone
Ux at open switch [Normal operation]:
1.6 V
Ux at closed switch [Alarm] :
1.1 V
Summary of Contents for SUPREMA Touch
Page 2: ...Manual SUPREMATouch Fire and Gas Warning Unit Order No 10126972 00...
Page 7: ...SUPREMATouch 6 Contents MSA US...
Page 8: ...User Instruction Manual SUPREMATouch Fire and Gas Warning Unit...
Page 104: ...Service and Maintenance Guide SUPREMATouch Fire and Gas Warning Unit...
Page 112: ...Installation and Start Up Manual SUPREMATouch Fire and Gas Warning Unit...
Page 151: ...SUPREMATouch 150 Installation MSA US Fig 79 MCP Module standard configuration...
Page 303: ...SUPREMA 302 Dimensions MSA GB 16 Dimensions 16 1 Rack...
Page 306: ...MSA AUER MSA Dimensions SUPREMA 305 GB MRO20 8 TS Module 1 3 2 69 90...
Page 307: ...SUPREMA 306 Dimensions MSA GB MRO20 16 TS Module 2 5 2 64 73 relay dependent 90...
Page 308: ...MSA AUER MSA Dimensions SUPREMA 307 GB MRC TS Module MGT 40 TS Module...
Page 309: ...SUPREMA 308 Dimensions MSA GB MHD TS Module MAT TS Module...