MSA AUER
MSA
SUPREMA Sensor Data Sheets
SUPREMA
295
US
Startup:
Presetting required —> before first calibration and when changing sensor
Presetting:
Connect digital voltmeter to MAI card jacks.
Bridge current setting —> 280 mA
Zero adjustment by zero gas —> Zero setting to
Ua = 400 … 450 mV
Sensitivity adjustment with measuring gas -> Measuring range level
Ua = 1950 ... 2100 mV
or by means of the value of the existing gas concentration
according to:
Ua (mV) = C (Span gas concentration in % of measuring range)/
100 * 1600 + 400
Warm-up
period:
15 min minutes for presetting,
2 hours for calibration
Function test: Span gas application via: test cap with 1,0 l/ min (Order-No.: 10049316)
Calibration:
Calibration procedure according to SUPREMA operation manual
For measuring components allowed, measuring ranges, lower alarm levels and
conditions for calibration see list of components (Order-No.: D0792420)
Possible other measuring components and measuring ranges on request.
Open or Short Circuit Fault Indication:
X=
Signal failure (FAIL-LED)
XX=
Alarm LED‘s, Signal exceeded, Signal failure (FAIL-LED)
XXX=
only alarms
XXXX=
no change of indication
Open-circuit
at the MAT
(TS) Module
Wire
-X1/1
Wire
-X1/2
Wire
-X1/3
Wire
-X1/4
Wire
-X1/5
Disconnect plug of
MAT (TS)
Failure
indication
XX
X
X
X
X
X
Open-circuit
at max. cable
length
Wire
-X1/1
Wire
-X1/2
Wire
-X1/3
Wire
-X1/4
Wire
-X1/5
Disconnect plug of
MAT (TS)
Failure
indication
XX
X
X
X
X
X
Short-circuit
at the MAT
(TS) Module
Wire
-X1/1/-
X1/2
Wire
-X1/1/-
X1/3
Wire
-X1/1/-
X1/4
Wire
-X1/1/-
X1/5
Wire
-X1/2/-
X1/3
Wire
-X1/2/-
X1/4
Wire
-X1/2/-
X1/5
Wire
-X1/3/-
X1/4
Wire
-X1/3/-
X1/5
Wire
-X1/4/-
X1/5
Failure
indication
X
XX
X
X
XX
XX
X
X
X
XX
Short-circuit
at max. cable
length
Wire
-X1/1/-
X1/2
Wire
-X1/1/-
X1/3
Wire
-X1/1/-
X1/4
Wire
-X1/1/-
X1/5
Wire
-X1/2/-
X1/3
Wire
-X1/2/-
X1/4
Wire
-X1/2/-
X1/5
Wire
-X1/3/-
X1/4
Wire
-X1/3/-
X1/5
Wire
-X1/4/-
X1/5
Failure
indication
XXXX
XX
X
X
XX
XX
X
X
X
XXXX
Summary of Contents for SUPREMA Touch
Page 2: ...Manual SUPREMATouch Fire and Gas Warning Unit Order No 10126972 00...
Page 7: ...SUPREMATouch 6 Contents MSA US...
Page 8: ...User Instruction Manual SUPREMATouch Fire and Gas Warning Unit...
Page 104: ...Service and Maintenance Guide SUPREMATouch Fire and Gas Warning Unit...
Page 112: ...Installation and Start Up Manual SUPREMATouch Fire and Gas Warning Unit...
Page 151: ...SUPREMATouch 150 Installation MSA US Fig 79 MCP Module standard configuration...
Page 303: ...SUPREMA 302 Dimensions MSA GB 16 Dimensions 16 1 Rack...
Page 306: ...MSA AUER MSA Dimensions SUPREMA 305 GB MRO20 8 TS Module 1 3 2 69 90...
Page 307: ...SUPREMA 306 Dimensions MSA GB MRO20 16 TS Module 2 5 2 64 73 relay dependent 90...
Page 308: ...MSA AUER MSA Dimensions SUPREMA 307 GB MRC TS Module MGT 40 TS Module...
Page 309: ...SUPREMA 308 Dimensions MSA GB MHD TS Module MAT TS Module...