EQUIPMENT PREPARATION
ELI 380 Configuration for All Users
Select
to enter the configuration menu. The following selections are available to all users. For more
detail, refer to
About:
Upon selection of the Configuration function icon, information is initially displayed for the ELI 380 serial
number, software version, DICOM enabled, WLAN enabled, LAN and WLAN MAC addresses.
Custom ID:
Selection of this function will initiate download of the Custom ID. The user is notified about success
of the download. Select
Done
to save changes and exit the menu, or
Cancel
prior to completion of download to
exit without saving.
Date/Time:
Use this function icon to manually enter date and time. The time zone can be set to your region as
well as daylight saving start and end times and dates. Date and time can also be synchronized with an available time
server. Select
Done
to save changes and exit the menu, or
Cancel
to exit without saving.
NOTE:
If the system goes through a hard reboot, the set Date/Time screen will appear at
start-up so the settings can be reset.
WAM/AM-XX:
Select to display the device version and to switch between use of AMxx or WAM. When
Switch
To WAM
function is selected, an additional
WAM Pairing
function becomes available and must be selected. Select
Done
to save changes and exit the menu.
NOTE:
The WAM must be paired to ELI 380 before operation. Refer to
WAM Wireless Acquisition Module
for instructions.
Network:
Select to evaluate network communication. WLAN or LAN information such as MAC Address, module
and radio firmware version, connection status, IP address and current signal strength is displayed. Select
Test
WLAN
or
Test LAN
to initiate a test with status information. Select
Done,
or
Cancel
during the WLAN or LAN
test to exit the menu.
Print:
Select to generate a printout of all ELI 380 configuration settings.
Options Code:
Select to present a field for options code entry. The code is obtained from Mortara Service
personnel to enable specific features. After the code is entered, select
Done
to save changes and exit the menu, or
Cancel
to exit without saving.
NOTE:
Selecting
Done
with improper code entry will cause an Invalid Code
notification. Select
Cancel
to exit.
Advanced:
Selection requires an administrative password and is explained in
Select
Cancel
to exit the menu.
Log On/Off:
When user authentication is enabled, select
Log On
to log on as a technician or administrator.
Select
Log Off
when done using the device or to use the device with a different account.
Select
at any time to return to the real-time ECG display.
33