CONNECTIVITY AND ECG TRANSMISSION
15- lead placement
In a 15-lead configuration, three combinations of additional lead wires are available. The following are the factory-
default lead sets:
1.
Pediatric: uses the right precordial leads V3R and V4R, and the posterior lead V7
2.
Posterior: uses the posterior leads V7, V8, and V9
3.
Right-sided: uses the right-sided leads V3R, V4R, and V6R
NOTE
: Attached to each 15 lead wire is a plug to be inserted into the AM15E. Each plug has a single label.
The label is either E2, E3 or E4. Insert each plug into the correspondingAM15E socket labeled E2, E3 and E4.
The E1 socket on the AM15E is not used. All 15 lead wire sets ordered from Mortara include a spacer. The
spacer prevents insertion of a lead wire plug into a socket. Please put the spacer into the E1 socket on the
AM15E to prevent insertion of a lead wire plug.
For limb leads and precordial leads V1 thru V6, follow the instructions for a standard 12-lead hookup as described
previously.
The factory-default Posterior and Right-sided lead sets can be renamed and redefined. The three leads can be
defined to be any of the following: V1, V2, V3, V4, V5, V6, V7, V8, V9, V3R, V4R, V5R, V6R, and V7R. The
lead sets are defined by selecting the
15 Leads Alt. Placement
button on the advanced settings page.
For posterior leads:
V7: left posterior axillary line, horizontal with V6.
V8: left midscapular line, horizontal with V7.
V9: left paraspinal line, horizontal with V8.