ECG DIRECTORY, MWL, AND THE PATIENT LIST
DIRECTORY
The ECG directory saves up to 500 individual ECG records. Records are automatically deleted through a first-in,
first-out process when the maximum number allowed has been reached.
To access the ECG directory, select
from the real-time display and then tap the
Directory
tab if not
displayed. Access requires the user to be logged in as a Technician or Administrator, or to know the device’s
Technician password depending on the ELI 380 is configured.
Sort the directory by
Name
,
ID
,
DOB
, or
Last Acquisition
by a tap on the column header. A second tap will sort
the list in reverse order.
The
Search
field can be used to locate patient records by entering free text for name, ID, DOB, or acquisition date.
The listing will refresh as characters are entered. Tap the
X
at the right of the search field to clear it.
Use the double arrows (
<<
or
>>
) at the right center of the screen to move page by page through the list of acquired
ECGs. The current/total number of pages is displayed between the double arrows.
An ECG with a Critical Test Result is flagged as such and easily identified in the Directory and Record lists (if CTR
configuration setting is on).
When a patient is selected from the directory, a subdirectory is presented showing all ECG records stored for that
patient with the acquisition date and time of each. Records that have been printed, transmitted, or deleted will be
indicated by an
X
in the appropriate column.
54