CONNECTIVITY AND ECG TRANSMISSION
Changing Best 10 or Last 10
1.
Select
from the real-time display to acquire an ECG.
2.
From acquired ECG view, touch anywhere on the red grid ECG to open the configuration screen.
3.
Select
Best10
or
Last10
.
4.
Select
OK
to save the selection and reformat, print, and display the ECG, or
Cancel
to exit without any
change.
NOTE:
If a single limb lead or two precordial lead fail condition occurs, the Best 10 feature is disabled
until the limb lead fail or precordial lead fail condition is resolved. Once the condition is corrected, the
Best 10 feature automatically becomes available.
NOTE:
Best10 function will automatically recognize signal dropout (e.g. with the WAM front
end) and will not use dropout in creating the Best10 result.
ECG Selection from Full Disclosure
In the real-time ECG view, the bottom of the screen displays a single lead of the 12-lead context view.
At any time during the ECG acquisition you may move the cursor over an area of interest and select the ECG data
for review and analysis.
Tap the lead label to display a menu allowing you to choose any other single lead.
Use
and
to page up or down through the full disclosure.
Tap
in this area of the window to provide a preview of the ECG on the main ECG panel for storage,
printing, transmission, or erasure.
50