CONFIGURATION SETTINGS
Daylight Saving:
Use the dropdown window to select
Yes
or
No
as observed within your region.
Time Zone:
Use the dropdown window to select the regional time zone. Use the up/down arrows to advance to
more selections.
Daylight Saving Start:
Enter the month, week, day, hour, and minute that daylight savings will begin for your
region.
Daylight Saving End:
Enter the month, week, day, hour, and minute that daylight savings will end for your
region.
Sync Date/Time:
Allows for time synchronization to occur only when the ELI 380 is interfaced via ELI Link to a
product that supports time synchronization. Select
Sync Date/Time
to synchronize with an available time server.
NOTE:
This device supports automatic time synchronization with a remote server. Inaccurate time/date
settings can result in ECGs marked with inaccurate time stamps. Confirm the accuracy of the synchronized
time prior to acquiring ECGs.
Select
Done
to save changes and exit this menu, or
Cancel
to exit without saving.
Configuration Menu: WAM/AMXX
Selecting
WAM/AMXX
allows the clinician to switch between WAM or AMxx patient interface cables. For pairing
of the WAM with the ELI 380, refer to
Configuring the WAM Wireless Acquisition Module
Configuration Menu: Network
Selecting
Network
establishes communication with your institution LAN or WLAN network and displays up to
five bars of signal strength. Presentation of the MAC Address, Firmware module, Radio Firmware, and IP address
connected to will also be presented. Refer to the
Connectivity and ECG Transmission
section for details.
Configuration Menu: Print
Selecting
will generate a printout of all ELI 380 configuration settings.
Configuration Menu: Options Code
Select
Options Code
to present a field for options code entry. The code is obtained from Mortara Service
personnel to enable specific features. After the code is entered, select
Done
to save changes and exit menu, or
Cancel
to exit without saving. Improper code entry will cause an
Invalid Code
notification.