9.
EQUIPMENT PREPARATION
Initial Startup
When powering the ELI 380 on for the first time, the device requires the user to set certain configurations prior to
obtaining an ECG:
1.
Custom ID (if applicable). Refer to the ELI Link user manual for instructions on how to set and change the
Custom ID.
2.
Date and time, including daylight savings start and end time selection.
3.
Configure the acquisition module.
4.
Advanced ECG and system settings. Refer to
for instructions.
NOTE
: The WAM, AM12, AM12M, and AM15E acquisition modules can be used with the ELI 380;
however, only one acquisition module can be used per ECG acquisition. To switch from one acquisition
module to the other prior to ECG acquisition, see the configuring instructions below.
NOTE:
When using a Wireless Acquisition Module (WAM) care must be taken to ensure that the expected
WAM module is paired with the cardiograph. It is recommended to use visual identifying indicators (like
color coded or matching labels) on the WAM and Cardiograph to avoid mismatch between units.
Configuring the AMxx Acquisition Module
Connect the AMxx to the ECG connector at the rear of the device and configure it at the ELI 380 before operation.
At the ELI 380 select
followed by
WAM/AM-XX
. Dependent on last saved setting, either AM12,
AM15E, or WAM is displayed with FPGA and UTK Firmware versions. Select
Switch to AM-XX
and then select
Done
to return to the Configuration screen.
Figure 7 ELI 380 AMxx Connection
AMxx USB Connection Port
Configuring the WAM Wireless Acquisition Module
When using the WAM for ECG acquisition, no physical connection is required. WAM must be selected and paired
to the ELI 380 before operation. Only one WAM can be configured to one ELI 380. The same WAM will remain
paired with the electrocardiograph for future use. A different WAM must be paired to the ELI 380 before use.
31