MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
Sinus 10s (150): 20.6 mGy-cm
T Bone 40s (600): 62.8 mGy-cm
SourceRay Source
Sinus 40s (600): 97.0 mGy-cm
Sinus 20s (300): 49.2 mGy-cm
Sinus 10s (150): 24.6 mGy-cm
T Bone 40s (600): 76.9 mGy-cm
Tube current modulation or
dose reduction technique
Pulsed Beam, Limited mA, Collimation
Total effective dose estimate
per acquisition
Sinus 20s (600): 0.17 mSv
Sinus 10s (300): 0.09 mSv
Sinus 10s (150: 0.04 mSv
T Bone 20s (600): 0.14 mSv
Prescribed rotation time
Sinus 20s (600): 20 seconds
Sinus 10s (300): 10 seconds
Sinus 10s (150): 10 seconds
T Bone 20s (600): 20 seconds
Sinus 40s (600): 40 seconds
Sinus 20s (300): 20 seconds
Sinus 10s (150): 10 seconds
T Bone 40s (600): 40 seconds
Increment
N/A
Detector collimation (mm)
Sinus: 189 mm
Temporal Bone: 105 mm
Slice thickness
Sinus: Selectable (0.4 mm minimum reconstructed)
Temporal Bone: Selectable (0.3 mm minimum reconstructed)
Pitch or table feed
N/A
Scan FOV
Sinus: 16 x 10.5 cm
Temporal Bone: 16 x 6 cm
Gantry tilt/angle
N/A
Kernel/filter
Sinus: Hann
Temporal Bone: Ramp