99
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
Secondary Dose Measurements Per Scan Level with Central Slice
Azimuthal Location
Radius (meter)
Dose Per Scan (millirad)
A
1.0
0.877
1.5
0.438
2.0
0.257
B
1.0
0.919
1.5
0.420
2.0
0.261
C
1.0
1.050
1.5
0.438
2.0
0.257
D
1.0
1.101
1.5
0.438
2.0
0.271
E
1.0
1.092
1.5
0.480
2.0
0.271
F
1.0
0.812
1.5
0.373
2.0
0.261
The secondary radiation doses were also recorded at 0.5 meters above and 0.5 meters below the
horizontal plane intersection the central slice at azimuthal locations A through F and at a radius of 1.0
meter. These measurements are reported in the following tables.