38
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
5.0 DAILY CALIBRATION
For best image quality, the MiniCAT™ system must be calibrated daily. Calibration should be
performed after the system has been powered on for at least one hour. It is recommended to leave
the MiniCAT™ system on overnight and calibrate each morning.
Caution:
A patient must not be in the scanner during Daily Calibration. X-rays will be
emitted during Daily Calibration.
Daily Calibration Procedure
1.
The area between the X-ray source and detector must be free of all objects.
Remove
the head holder, phantom holder and any phantom that might be present before continuing.
2a. Select the
Daily Calibration
option from the
Tools
menu.
2b. Alternatively, you can push the F9/CAL key.
3. A message will appear to confirm that all objects have been removed from the field of view.
Click
OK
to proceed.
4. A message will appear to start the calibration process. Click
Calibrate
to begin.
5. A message will appear asking you if it is
OK
to turn on the X-ray source. Ensure that you are
behind the radiation barrier and click
OK
.
Caution:
It is the user's responsibility to ensure that proper radiation safety procedures
are followed before X-rays are emitted.
6. Remain behind the radiation barrier until a message appears indicating the calibration is complete.
The calibration procedure will take approximately 5 minutes to complete.