103
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
(c)(2)(i)(c)(iv)
Dose profile for each selectable nominal
tomographic section thickness*.
Please see results above
for CTDI values and
Section 10.2 for slice
sensitivity profiles.
N/A
(c)(2)(i)(c)(v)
A statement of the maximum deviation
from these results
All protocols
30%
*The dose profile for each selectable nominal tomographic section thickness is not available because
the patient position is fixed throughout the acquisition. The dose profile does not vary across the
field of view because of the cone beam geometry. The nominal tomographic section thickness is
fixed for each protocol and is not selectable prior to acquisition.
Patient Effective Dose Estimates
Imaging Protocol
Total Effective Dose
Estimate (mSv)
Sinus 20s (600)
0.17
Sinus 10s (300)
0.09
Sinus 10s (150)
0.04
T Bone 20s (600)
0.13
Scout View
< 0.01
Legacy Protocols (ISI Source)
The CT conditions of operation used to provide dose and imaging performance information below
are as follows:
Sinus 40s (600): 120 kVp, 5 mA, 16 ms pulses, 600 pulses (48 mAs), 10.6 mm Al equivalent fixed
filtration, 0.542 ± 0.048 mm tomographic section thickness.
Sinus 20s (300): 120 kVp, 5 mA, 16 ms pulses, 300 pulses (24 mAs), 10.6 mm Al equivalent fixed
filtration, 0.547 ± 0.071 mm tomographic section thickness.
Sinus 10s (150): 120 kVp, 5 mA, 16 ms pulses, 150 pulses (12 mAs), 10.6 mm Al equivalent fixed
filtration, 0.536 ± 0.130 mm tomographic section thickness.
T Bone 40s (600): 125 kVp, 7 mA, 18 ms pulses, 600 pulses (75.6 mAs), 10.6 mm Al equivalent
fixed filtration, 0.532 ± 0.043 mm tomographic section thickness.
The nominal tomographic section thickness is fixed for each protocol and is not selectable prior to