105
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
Note:
The operator of the MiniCAT™ must be able to make visual and audio contact with the
patient being scanned while behind radiation barrier.
Power
Requirement
Japan
Japan
Europe
USA
Line Voltage
~200V
~100V
~220-240V
~115V
Line Frequency
50-60 Hz
50-60 Hz
50-60 Hz
50-60 Hz
Line Current
r.m.s (standby)
(@ line voltage)
2.5A
5A
2.5A
5A
Line Current
r.m.s (max)
Measured @
Line voltage,
load 125 kVp,
58.8 mAs
5A
10 A
5 A
10 A
Phase
Single
Single
Single
Single
Main Circuit
Breaker
5A
10A
5A
10A
MiniCAT input voltage regulation range (measured at UPS output): 100 x (V
NO LOAD
–V
LOADED
)/V
LOADED
NB: unit must be used with provided line conditioner (UPS); Do not initiate a scan without AC power to
the UPS.
* For immunity to electrical noise, a dedicated AC power line with a separate circuit breaker, dedicated
neutral and ground is recommended to install a MiniCAT.
A dedicated line is required if the above
voltage regulation requirem ent at the UPS output cannot be m et.
Line frequency range 47 Hz – 63 Hz
Note:
A dedicated electrical wall outlet with a separate circuit breaker and ground is
recommended to protect against electrical noise.