19
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
2.6 Quick Reference Guide
Note:
This page is a brief overview of the MiniCAT™. Please attend MiniCAT™ training with a
qualified instructor, and read the complete MiniCAT™ User Manual before operating the
equipment.
PREPARATION
1. Perform a Daily Calibration
For best results, scanner should be powered on for at least one hour before calibration
Note:
Refer to Section 3.2 for System Startup instructions.
Click
Daily Calibration
from the
Tools
menu (or press the F9/CAL key)
Caution:
X-ray radiation emitted. Follow radiation safety protocols.
2. Enter Patient Information
Enter patient information and click
Save
SCANNING
1.
Select Patient
From the
Patient Database
, locate and highlight the patient record
Click
Acquire New
and follow the on-screen prompts
2. Select Procedure
Create a study title, select the procedure and choose the ordering physician
3. Seat and Position Patient
Have patient remove jewelry, glasses, hair accessories, etc. from patient's head
Adjust height of patient chair until ear is positioned as shown
Use alignment light to position patient based on imaging protocol (see below)
Caution:
Laser light emitted. Do not stare into the beam.