
114
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
1. Scan the QC Phantom with the QC module placed on the central slice, the Front label facing
forward, and the entire module placed centrally in the field of view. Scan the phantom with the
appropriate protocol(s) specified in the respective QC test.
11.1 Laser Accuracy
To confirm alignment light accuracy:
1. Press and hold the
Alignment Light
button on the controller to turn on laser.
Press button once to turn on the laser for a few seconds.
Press and hold for three seconds to turn on the laser for five minutes.
Side lasers should align with the alignment stickers on perimeter of X-ray detector.