116
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
Air
-1000 to -650 HU
Teflon
>550 HU
Acrylic 50 to 250 HU
LDPE
-200 to 0 HU
11.3 Spatial Resolution
Equipment and Scan:
Quality Control Phantom, Resolution QC scan
High Contrast Procedure:
1. Using the Resolution QC scan, position the Quality Control phantom at the central slice.