101
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
fixed filtration, 0.547 ± 0.071 mm tomographic section thickness.
Sinus 10s (150): 120 kVp, 7 mA, 11.5 ms pulses, 150 pulses (12.08 mAs), 11.7 mm Al equivalent
fixed filtration, 0.536 ± 0.130 mm tomographic section thickness.
T Bone 20s (600): 125 kVp, 7 mA, 14.0 ms pulses, 600 pulses (58.8 mAs), 11.7 mm Al equivalent
fixed filtration, 0.532 ± 0.043 mm tomographic section thickness.
The nominal tomographic section thickness is fixed for each protocol and is not selectable prior to
acquisition.
Dose Measurements - SourceRay Source
Protocol
CTDI
VOL
(mGy)
DLP (mGy-cm)
Sinus 20s (600)
Sinus 40s (600)
5.1
97.0
Sinus 10s (300)
Sinus 20s (300)
2.6
49.2
Sinus 10s (150)
1.3
24.6
T Bone 20s (600)
T Bone 40s (600)
7.3
76.9
21CFR1020.33
Subsection
Description
Protocol
(SourceRay
Source)
Result
(c)(2)(i)(a)
CTDI at center for typical factors
Sinus 20s (600)
Sinus 40s (600)
5.4 mGy
CTDI at center for typical factors
T Bone 20s (600)
7.7 mGy