126
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
27. Record the Teflon mean value in the CT
x
section of the User QC Spreadsheet.
28. Record the coordinates (X, Y) of center of the Teflon region in the (X,Y) section of the User
QC spreadsheet. (NOTE: The coordinates are located in the upper-right region of the MPR
window.)
29. Scroll axially in the superior direction about 3 cm to the Water Module of the QC Phantom.
Draw a 65-80 mm
2
circular region at (approximately) the same (X,Y) location. Record the
mean value in this region in the CT
w
Section of the User QC Spreadsheet.
30. Repeat Steps 26-29 for Sinus 10s (300), Sinus 10s (150), and T Bone 20s (600).
Tolerance:
Sinus 20s (600): 0.00023 to 0.00034
Sinus 10s (300): 0.00021 to 0.00041