89
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
Tomographic Section Thickness
The following table gives the tomographic section thickness for each imaging protocol with maximum
deviations.
Protocol Name(s)
Tomographic Section
Thickness (mm)
Sinus 20s (600), Sinus 40s (600)
0.542 ± 0.048
Sinus 10s (300), Sinus 20s (300)
0.547 ± 0.071
Sinus 10s (150)
0.536 ± 0.130
T Bone 20s (600), T Bone 40s (600)
0.532 ± 0.043
Measuring Nominal Tomographic Section Thickness is obtained by finding the full-width half-
maximum (FWHM) resolution of the Slice Sensitivity Profile (SSP). The SSP is obtained by
scanning a small, sub-voxel bead in the QC module. However, the nominal tomographic section
thickness is fixed for each protocol and is not selectable prior to acquisition.
Slice Sensitivity Profiles