A-2
A.3 Operating Range
Temperature
During operation
10 to 40
℃
During storage (empty)
-20 to 60
℃
During transport (empty)
-20 to 60
℃
Humidity
During operation
15 to 95%, non-condensing
During storage
10 to 95%, non-condensing
Atmospheric pressure
During operation and shut-down (filled,
control dial at “0” position)
70 to 106 kPa
During storage (empty)
50 to 120 kPa
Concentration range
Isoflurane
0 to 6%
Sevoflurane
0 to 8%
Enflurane
0 to 5%
Halothane
0 to 5%
A.4 Performance Specifications
Range of concentration accuracy (maximum value always applies)
Operating environment
15 to 35
℃
or 0.2 to 10 L/min
10 to 15
℃
or 35 to 40
℃
or 10 to
15 L/min
Set concentration
≤
6%
±0.20 vol.% or ±20% rel.,
whichever is greater
+0.30/-0.20 vol.% or +25/-20%
rel., whichever is greater
Set concentration > 6%
±0.25 vol.% or ±20% rel.,
whichever is greater
+0.35/-0.25 vol.% or +30/-20%
rel., whichever is greater
Maximum angle of tilt
Alone, freestanding
10
º
During operation (fixed position)
30
º
Pressure difference
Difference between pressure range and ambient pressure on the
vaporizer outlet
-10 to 10 kPa
Summary of Contents for V60
Page 1: ...V60 Anesthetic Vaporizer Operator s Manual ...
Page 2: ......
Page 9: ...3 C Symbols and Terminology C 1 C 1 Symbols C 1 C 2 Terminology C 1 ...
Page 10: ...4 FOR YOUR NOTES ...
Page 17: ...2 3 2 3 2 Rear View 1 Handle for locking lever 2 Interlock system 3 Locking pin ...
Page 48: ...5 4 FOR YOUR NOTES ...
Page 56: ...6 8 FOR YOUR NOTES ...
Page 66: ...9 4 FOR YOUR NOTES ...
Page 79: ...11 11 0 1 2 3 4 5 6 0 1 2 3 4 5 6 G C B A F Concentration delivered vol Enflurane Time h ...
Page 80: ...11 12 ...
Page 84: ...A 4 FOR YOUR NOTES ...
Page 86: ...B 2 FOR YOUR NOTES ...
Page 89: ......
Page 90: ...P N 046 001002 00 8 0 ...