4-7
CAUTION
z
It is necessary to wait at least 5 seconds after setting the control dial to the 0 position
before opening the vaporizer. This allows the pressure to balance and prevents fresh
gas and anesthetic agent vapor froming escaping from the vaporizer.
2. Select the correct filling adapter and anesthetic agent bottle. Screw the filling adapter firmly into
the anesthetic agent bottle. Before use, check that the color marks and names/symbols of anesthetic
agent on the filling adapter, anesthetic agent bottle and vaporizer must correspond to the anesthetic
agent used.
WARNING
z
Do not use a damaged filling adapter or an anesthetic agent bottle without collar. If a
bottle without collar is used, specific filling adapter cannot be identified, underlying the
risk of filling a wrong anesthetic agent.
CAUTION
z
If the connection between the filling adapter and anesthetic agent bottle is not
leak-tight, anesthetic agent may escape.
NOTE
z
If a new anesthetic agent bottle is partly empty, there may be a leak.
Filling adapter
Anesthetic agent bottle
Bottle neck collar
Summary of Contents for V60
Page 1: ...V60 Anesthetic Vaporizer Operator s Manual ...
Page 2: ......
Page 9: ...3 C Symbols and Terminology C 1 C 1 Symbols C 1 C 2 Terminology C 1 ...
Page 10: ...4 FOR YOUR NOTES ...
Page 17: ...2 3 2 3 2 Rear View 1 Handle for locking lever 2 Interlock system 3 Locking pin ...
Page 48: ...5 4 FOR YOUR NOTES ...
Page 56: ...6 8 FOR YOUR NOTES ...
Page 66: ...9 4 FOR YOUR NOTES ...
Page 79: ...11 11 0 1 2 3 4 5 6 0 1 2 3 4 5 6 G C B A F Concentration delivered vol Enflurane Time h ...
Page 80: ...11 12 ...
Page 84: ...A 4 FOR YOUR NOTES ...
Page 86: ...B 2 FOR YOUR NOTES ...
Page 89: ......
Page 90: ...P N 046 001002 00 8 0 ...