11.4 Influence of Flow
Within the specified flow range, the concentration delivered by the vaporizer is only slightly dependent
on the fresh gas flow.
In case of high fresh gas flow or high concentration, full compensation is not made for the cooling of
the anesthetic agent because total saturation of the gas flowing through the liquid vaporizing system
does not occur and the output concentration is reduced slightly (see
11.8 Influence of Running Time
).
The diagrams show the influence of flow on the concentration delivered after 1 minute at 22
℃
, 101.3
kPa when operating with Air.