4.3.3 Quik-Fil System
Place the vaporizer upright or suspend so that all the anesthetic agent can drain out.
1. Turn the control dial clockwise back to “0” position until the “0” button pops up.
2. Remove the cap from an empty Sevoflurane anesthetic agent bottle. Screw the Quik-Fil
adapter firmly into the anesthetic agent bottle. Insert the drainage funnel. Depress the
drainage funnel and rotate the bottle simultaneously until it cannot move.
CAUTION
z
If the connection between the filling adapter and anesthetic agent bottle is not
leak-tight, anesthetic agent may escape, endangering health.
3. Insert the bottle equipped with Quik-Fil drainage funnel into the slot at the bottom of the
vaporizer.
Depress
Rotate
Drainage funnel
Anesthetic agent bottle