4-3
CAUTION
z
Make sure that the drainage knob is closed when filling the vaporizer as anesthetic
agent may escape from the drainage outlet if it is not closed.
z
Keep the vaporizer upright or hanging vertical while it is being filled. If it is at an angle
it can be overfilled which may lead to concentrations which are too high or too low.
z
During disconnection of Key Filler and Quik-Fil filling adaptor from the vaporizer and
the bottle adaptor from the bottle, small amounts of anesthetic agent will escape to the
environment.
4.2.1 Pour Fill System
The filling steps of V60 Isoflurane Pour Fill Vaporizer and V60 Sevoflurane Pour Fill Vaporizer are the
same.
If the vaporizer is connected to the anesthetic delivery system, fresh gas flow can remain as set.
1. Turn the control dial clockwise back to “0” position until the “0” button pops up.
WARNING
z
Significant quantities of anesthetic agent vapor may escape if the control dial does not
return to “0” position.
Turn the control dial back to
“0” position
Summary of Contents for V60
Page 1: ...V60 Anesthetic Vaporizer Operator s Manual ...
Page 2: ......
Page 9: ...3 C Symbols and Terminology C 1 C 1 Symbols C 1 C 2 Terminology C 1 ...
Page 10: ...4 FOR YOUR NOTES ...
Page 17: ...2 3 2 3 2 Rear View 1 Handle for locking lever 2 Interlock system 3 Locking pin ...
Page 48: ...5 4 FOR YOUR NOTES ...
Page 56: ...6 8 FOR YOUR NOTES ...
Page 66: ...9 4 FOR YOUR NOTES ...
Page 79: ...11 11 0 1 2 3 4 5 6 0 1 2 3 4 5 6 G C B A F Concentration delivered vol Enflurane Time h ...
Page 80: ...11 12 ...
Page 84: ...A 4 FOR YOUR NOTES ...
Page 86: ...B 2 FOR YOUR NOTES ...
Page 89: ......
Page 90: ...P N 046 001002 00 8 0 ...