6-4
When the anesthetic delivery system has three and more than three groups of plug-in
connectors, check if there is an interlock function between nonadjacent connectors.
Otherwise, vaporizers are recommended to be connected right next to each other.
WARNING
z
The plug-in adapter must be level and stable on the sealing rings. Otherwise, there may
be a loss of fresh gas, leaks, excessively low output concentrations or the interlock
device may jam. To solve this problem, disconnect the vaporizer first (see 6.4
Disconnecting the Vaporizer) and check the positions of locking lever and vaporizer
manifold of the anesthetic delivery system. Then re-connect the vaporizer.
6.2 Adjusting the Concentration of Anesthetic Agent
WARNING
z
Before operation, check that the control dial turns normally.
z
Do not use the vaporizer tilted for an angle of more than 30°(fixed position). Risk of
incorrect output concentration or escape of anesthetic agent may result otherwise.
1. Set the flow of fresh gas on the anesthetic delivery system.
2. Press the “0” button.
3. Turn the control dial counter clockwise to the required concentration of anesthetic agent.
NOTE
z
Take care when lowering the Vaporizer onto the plug-in connector
Summary of Contents for V60
Page 1: ...V60 Anesthetic Vaporizer Operator s Manual ...
Page 2: ......
Page 9: ...3 C Symbols and Terminology C 1 C 1 Symbols C 1 C 2 Terminology C 1 ...
Page 10: ...4 FOR YOUR NOTES ...
Page 17: ...2 3 2 3 2 Rear View 1 Handle for locking lever 2 Interlock system 3 Locking pin ...
Page 48: ...5 4 FOR YOUR NOTES ...
Page 56: ...6 8 FOR YOUR NOTES ...
Page 66: ...9 4 FOR YOUR NOTES ...
Page 79: ...11 11 0 1 2 3 4 5 6 0 1 2 3 4 5 6 G C B A F Concentration delivered vol Enflurane Time h ...
Page 80: ...11 12 ...
Page 84: ...A 4 FOR YOUR NOTES ...
Page 86: ...B 2 FOR YOUR NOTES ...
Page 89: ......
Page 90: ...P N 046 001002 00 8 0 ...